A novel electromagnetic characterization of Implantable Cardiac Medical Devices (ICMDs) is presented in this paper. It has been developed in order to determine the immunity level of ICMDs against Radio-Frequency (RF) interference and consists of several Sensitivity Tests, which have been performed in accordance with EMC international standards. The tests have been carried out in an anechoic RF chamber, which guarantees an appropriate management of noise signal injection, shielding against external RF noise sources at the same time. In this way, the operating thresholds of an Implantable Cardiac Defibrillator (ICD) have been achieved at different heartbeat frequency, with and without high-frequency noise. As a result, the comparison between the operating thresholds highlight the electromagnetic susceptibility of the ICD under test.

EMC Characterization of Implantable Cardiac Medical Devices in an Anechoic Chamber

SPANO, IVAN LUIGI;SULIS, SARA;SERPI, ALESSANDRO;MARONGIU, IGNAZIO;GATTO, GIANLUCA
2014-01-01

Abstract

A novel electromagnetic characterization of Implantable Cardiac Medical Devices (ICMDs) is presented in this paper. It has been developed in order to determine the immunity level of ICMDs against Radio-Frequency (RF) interference and consists of several Sensitivity Tests, which have been performed in accordance with EMC international standards. The tests have been carried out in an anechoic RF chamber, which guarantees an appropriate management of noise signal injection, shielding against external RF noise sources at the same time. In this way, the operating thresholds of an Implantable Cardiac Defibrillator (ICD) have been achieved at different heartbeat frequency, with and without high-frequency noise. As a result, the comparison between the operating thresholds highlight the electromagnetic susceptibility of the ICD under test.
2014
978-1-4799-3226-9
Electromagnetic compatibility, Immunity testing, Implantable Cardiac Defibrillator
File in questo prodotto:
File Dimensione Formato  
P41#2014#EMC-EU#IEEE_Paper.pdf

Solo gestori archivio

Descrizione: EMC Characterization of Implantable Cardiac Medical Devices in an Anechoic Chamber
Tipologia: versione editoriale
Dimensione 1.78 MB
Formato Adobe PDF
1.78 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
P41#2014#EMC-EU#PostPrint.pdf

Solo gestori archivio

Descrizione: EMC Characterization of Implantable Cardiac Medical Devices in an Anechoic Chamber (Post Print)
Tipologia: versione post-print
Dimensione 1.76 MB
Formato Adobe PDF
1.76 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11584/108903
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 4
social impact