Abstract PURPOSE: Our purpose was to investigate the multi-detector-row CT angiography (MDCTA) application in endoleak detection. METHODS: Fifty-nine patients that underwent endovascular aneurysm repair (EVAR) of an infrarenal abdominal aortic aneurysm with endoluminal stent graft were retrospectively studied. MDCTA scans were obtained after administration of 110-130 mL of contrast material using a 4-6 mL/sec flow rate. We made unenhanced, arterial (15-20 sec) and delayed (100 sec) acquisitions. For each patient four MDCTA datasets (pre-EVAR, 1, 6, and 12 months follow-up) were obtained. Each examination was studied by two observers. Kappa value was calculated in order to evaluate inter-observer agreement. RESULTS: Twenty-one endoleaks were detected in eighteen patients. Fifteen and eighteen endoleaks were detected by using biphasic arterial CT and biphasic delayed CT respectively (sensitivity of 71.4% and 85.7% respectively). We observed a positive correlation between the presence of endoleak and the increasement in aneurysm size. Interobserver agreement was 88.1% and kappa value was 0.685. CONCLUSIONS: Biphasic CT for endoleak detection is significantly superior to arterial and delayed phases respectively. We observed a good interobserver agreement.

Study of endoleaks after endovascular repair by using MDCTA

SABA, LUCA;SANFILIPPO, ROBERTO;MONTISCI, ROBERTO;
2010-01-01

Abstract

Abstract PURPOSE: Our purpose was to investigate the multi-detector-row CT angiography (MDCTA) application in endoleak detection. METHODS: Fifty-nine patients that underwent endovascular aneurysm repair (EVAR) of an infrarenal abdominal aortic aneurysm with endoluminal stent graft were retrospectively studied. MDCTA scans were obtained after administration of 110-130 mL of contrast material using a 4-6 mL/sec flow rate. We made unenhanced, arterial (15-20 sec) and delayed (100 sec) acquisitions. For each patient four MDCTA datasets (pre-EVAR, 1, 6, and 12 months follow-up) were obtained. Each examination was studied by two observers. Kappa value was calculated in order to evaluate inter-observer agreement. RESULTS: Twenty-one endoleaks were detected in eighteen patients. Fifteen and eighteen endoleaks were detected by using biphasic arterial CT and biphasic delayed CT respectively (sensitivity of 71.4% and 85.7% respectively). We observed a positive correlation between the presence of endoleak and the increasement in aneurysm size. Interobserver agreement was 88.1% and kappa value was 0.685. CONCLUSIONS: Biphasic CT for endoleak detection is significantly superior to arterial and delayed phases respectively. We observed a good interobserver agreement.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11584/101984
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact