The main purpose of the present study was to assess the effects of foot drop stimulators (FDS) in individuals with stroke by means of spatio-temporal and step-to-step symmetry, harmonic ratio (HR), parameters obtained from trunk accelerations acquired using a wearable inertial sensor. Thirty-two patients (age: 56.84 ± 9.10 years; 68.8% male) underwent an instrumental gait analysis, performed using a wearable inertial sensor before and a day after the 10-session treatment (PRE and POST sessions). The treatment consisted of 10 sessions of 20 min of walking on a treadmill while using the FDS device. The spatio-temporal parameters and the HR in the anteroposterior (AP), vertical (V), and mediolateral (ML) directions were computed from trunk acceleration data. The results showed that time had a significant effect on the spatio-temporal parameters; in particular, a significant increase in gait speed was detected. Regarding the HRs, the HR in the ML direction was found to have significantly increased (+20%), while those in the AP and V directions decreased (approximately 13%). Even if further studies are necessary, from these results, the HR seems to provide additional information on gait patterns with respect to the traditional spatio-temporal parameters, advancing the assessment of the effects of FDS devices in stroke patients.

Functional electrical stimulation for foot drop in post-stroke people: quantitative effects on step-to-step symmetry of gait using a wearable inertial sensor

Pau, Massimiliano
Formal Analysis
;
Leban, Bruno
Software
;
Porta, Micaela
Formal Analysis
;
2021-01-01

Abstract

The main purpose of the present study was to assess the effects of foot drop stimulators (FDS) in individuals with stroke by means of spatio-temporal and step-to-step symmetry, harmonic ratio (HR), parameters obtained from trunk accelerations acquired using a wearable inertial sensor. Thirty-two patients (age: 56.84 ± 9.10 years; 68.8% male) underwent an instrumental gait analysis, performed using a wearable inertial sensor before and a day after the 10-session treatment (PRE and POST sessions). The treatment consisted of 10 sessions of 20 min of walking on a treadmill while using the FDS device. The spatio-temporal parameters and the HR in the anteroposterior (AP), vertical (V), and mediolateral (ML) directions were computed from trunk acceleration data. The results showed that time had a significant effect on the spatio-temporal parameters; in particular, a significant increase in gait speed was detected. Regarding the HRs, the HR in the ML direction was found to have significantly increased (+20%), while those in the AP and V directions decreased (approximately 13%). Even if further studies are necessary, from these results, the HR seems to provide additional information on gait patterns with respect to the traditional spatio-temporal parameters, advancing the assessment of the effects of FDS devices in stroke patients.
2021
foot drop stimulation; gait; symmetry; stroke; inertial measurement sensor
File in questo prodotto:
File Dimensione Formato  
sensors-21-00921.pdf

accesso aperto

Tipologia: versione editoriale
Dimensione 505.44 kB
Formato Adobe PDF
505.44 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11584/306457
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 12
social impact