Purpose: To report a case of fully dislocated XEN Gel Implant device into the anterior chamber, 18 months after its implantation and its impact on endothelial cell density. Result: A 75-year-old man with a history of multiple glaucoma surgeries during the last three years presented with an entirely dislocated XEN Gel implant into the anterior chamber. An endothelial cell density reduction was observed over 18 months. Thus, the implant was removed. In the short-term after surgery, the endothelial cell loss rate reduced. Conclusion: XEN Gel Implant can dislocate into the anterior chamber, increasing endothelial cell loss in an eye already underwent multiple surgical procedures with subsequent adverse events. Therefore, if the implant seems displaced, the endothelial cell density should be monitored, and the length of the free-tube segment within the anterior chamber should be measured, during the patient’s follow-up.

Endothelial cell density change in fully dislocated XEN Gel IMPLANT after trabeculectomy: A case report

Tatti F.;Peiretti E.;Figus M.
2021-01-01

Abstract

Purpose: To report a case of fully dislocated XEN Gel Implant device into the anterior chamber, 18 months after its implantation and its impact on endothelial cell density. Result: A 75-year-old man with a history of multiple glaucoma surgeries during the last three years presented with an entirely dislocated XEN Gel implant into the anterior chamber. An endothelial cell density reduction was observed over 18 months. Thus, the implant was removed. In the short-term after surgery, the endothelial cell loss rate reduced. Conclusion: XEN Gel Implant can dislocate into the anterior chamber, increasing endothelial cell loss in an eye already underwent multiple surgical procedures with subsequent adverse events. Therefore, if the implant seems displaced, the endothelial cell density should be monitored, and the length of the free-tube segment within the anterior chamber should be measured, during the patient’s follow-up.
2021
dislocation
endothelial cell count
glaucoma
MIGS
XEN gel implant
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11584/316981
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 0
social impact