Ultrasound cyclo plasty (UCP) is a recently developed surgical technique for glaucoma allowing a selective and controlled coagulation of the ciliary body. We herein investigated the long-term efficacy and safety of UCP for the treatment of glaucoma. This prospective study included patients with primary and secondary glaucoma. All surgeries were performed using the EyeOP1 device (Eye Tech Care, Rillieux-la-Pape, France). Sixty-six patients were included, and 60 completed regularly the 2-year follow-up. Preoperative IOP was 28.5 ± 9.6 mmHg and significantly decreased to 17.0 ± 5.4 at 2 years (p < 0.001). The daily number of both hypotensive eye drops and acetazolamide tablets decreased significantly (respectively, from 2.6 ± 1.1 to 1.7 ± 1.2 and from 0.7 ± 0.8 to 0.2 ± 0.5; both p < 0.001). At 2 years, 68.1% of patients met the definition of qualified success (IOP < 21 mmHg regardless of glaucoma medications) and 10.3% of patients met the definition of complete success (IOP < 21 mmHg without glaucoma medications). No major intra- or postoperative complications occurred; however, 15 eyes required additional glaucoma surgery. These results suggest that UCP is an effective and safe procedure to reduce IOP in glaucoma patients through a 2-year follow-up period.

A 2-year prospective multicenter study of ultrasound cyclo plasty for glaucoma

Giannaccare G.
Primo
;
2021-01-01

Abstract

Ultrasound cyclo plasty (UCP) is a recently developed surgical technique for glaucoma allowing a selective and controlled coagulation of the ciliary body. We herein investigated the long-term efficacy and safety of UCP for the treatment of glaucoma. This prospective study included patients with primary and secondary glaucoma. All surgeries were performed using the EyeOP1 device (Eye Tech Care, Rillieux-la-Pape, France). Sixty-six patients were included, and 60 completed regularly the 2-year follow-up. Preoperative IOP was 28.5 ± 9.6 mmHg and significantly decreased to 17.0 ± 5.4 at 2 years (p < 0.001). The daily number of both hypotensive eye drops and acetazolamide tablets decreased significantly (respectively, from 2.6 ± 1.1 to 1.7 ± 1.2 and from 0.7 ± 0.8 to 0.2 ± 0.5; both p < 0.001). At 2 years, 68.1% of patients met the definition of qualified success (IOP < 21 mmHg regardless of glaucoma medications) and 10.3% of patients met the definition of complete success (IOP < 21 mmHg without glaucoma medications). No major intra- or postoperative complications occurred; however, 15 eyes required additional glaucoma surgery. These results suggest that UCP is an effective and safe procedure to reduce IOP in glaucoma patients through a 2-year follow-up period.
File in questo prodotto:
File Dimensione Formato  
41598_2021_Article_92233.pdf

accesso aperto

Tipologia: versione editoriale (VoR)
Dimensione 958.33 kB
Formato Adobe PDF
958.33 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11584/422927
Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 9
social impact